Repeatability of a Dual-Scheimpflug Placido Disc Corneal Tomographer/Topographer in Eyes with Keratoconus
Poulsen, A.; Jang, D.; Khan, M.; Al-Mohtaseb, Z. N.; Chen, M.; Zhan, X.; Banerjee, K.; Scott, I. U.; Pantanelli, S. M.
Show abstract
PurposeTo investigate the repeatability of a combined Dual-Scheimpflug placido disc corneal topographer (Zeimer Galilei G4) with respect to keratometric indices used to monitor progression of keratoconus (KCN). MethodsPatients with KCN were prospectively enrolled. For each eye lacking history of corneal surgery, 5 measurements were taken in succession. Eyes in which 3 or more measurements could be obtained (defined by the devices 4 image quality metrics) were included in the analysis. The repeatability limits (RL) and interclass correlation coefficients (ICC) were calculated for various parameters. Results32 eyes from 25 patients met all image quality metrics, and 54 eyes from 38 patients met at least 3/4 criteria (all except the placido image quality metric). RLs for key parameters when 4/4 or [≥]3/4 image quality metrics were met included: 0.37 and 0.77 diopters (D) for steep simulated keratometry, 0.79 and 1.65 D for maximum keratometry, 13.80 and 13.88 degrees for astigmatism axis, 0.64 and 0.56 m for vertical coma magnitude, and 3.76 and 3.84 m for thinnest pachymetry, respectively. The ICCs for all parameters were excellent [above 0.87 except for spherical aberration (0.77)]. ConclusionsThe dual-Scheimpflug placido disc corneal topographer is highly repeatable in quantifying parameters used in monitoring KCN. Excellent placido images are difficult to capture in eyes with KCN, but when available, increase the reliability of the measurements. The RLs may be especially helpful in detecting progression in mild KCN when interventions such as corneal cross-linking or intrastromal corneal ring segments are most beneficial.
Matching journals
The top 2 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Advanced Retinal Imaging and Ocular Parameters of the Rhesus Macaque Eye 95%
- Visible light optical coherence tomography of peripapillary retinal nerve fiber layer reflectivity in glaucoma 95%
- Relating Standardized Automated Perimetry Performed with Stimulus Sizes III and V in Eyes With Field Loss due to Glaucoma and NAION 95%
Similar papers in this journal
- Factors predicting the visual outcome of intracorneal ring segment for keratoconus 97%
- The evaluation of a web‐based tool for measuring the uncorrected visual acuity and refractive error in keratoconus eyes: a prospective open‐label method comparison study 96%
- Automated 360-degree goniophotography with the NIDEK Gonioscope GS-1 for glaucoma 96%
Similar papers in this journal
- Automated Expert-level Scleral Spur Detection and Quantitative Biometric Analysis on the ANTERION Anterior Segment OCT System 96%
- Evaluation of the Nallasamy Formula: A Stacking Ensemble Machine Learning Method for Refraction Prediction in Cataract Surgery 95%
- Optic nerve tortuosity and displacements during horizontal eye movements in healthy and highly myopic subjects 95%
Similar papers in this journal
- Iris Morphological and Biomechanical Factors Influencing Angle Closure During Pupil Dilation 94%
- Inner limiting Membrane Peel Extends In vivo Calcium Imaging of Retinal Ganglion Cell Activity Beyond the Fovea in Non-Human Primate 92%
- Proposing a methodology for axon-centric analysis of IOP-induced mechanical insult 92%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.